Cavity-Length Tuning and Angular Color Shift in Strong-Microcavity Top-Emitting OLEDs

Dahal et al. (2021): angle-resolved peak dispersion and quality factor of top-emitting OLEDs at the λ/2, λ, and 3λ/2 resonance orders
AuthorLuke Cole

Characterization of higher harmonic modes in Fabry–Pérot microcavity organic light emitting diodes

Authors: Ekraj Dahal, David Allemeier, Benjamin Isenhart, Karen Cianciulli, and Matthew S. White

Journal: Scientific Reports 11, 8456 (2021) · Comparison target: Figure 4

A top-emitting OLED can sweep its normal-direction peak across the entire visible range without changing the emitter, simply by making the organic stack thicker. Tilt the same device to 60° and the peak retreats tens of nanometres toward the blue.

This case builds three devices at the λ/2, λ, and 3λ/2 resonance orders, runs Intensity once, and reads the angle-resolved emission maps of paper Figure 4 together with the angular dispersion and quality factor.

Dahal Figure 4 — angle-resolved electroluminescence of the λ/2, λ, and 3λ/2 resonance-order devices; the left column is the authors' own TMM simulation and the right column is measurement.Dahal et al., Scientific Reports 11, 8456 (2021), Figure 4CC BY

The three rows are the three resonance orders. Every bright band bends toward the blue with observation angle, and the λ/2 row splits into two branches beyond 40°.

Background

Both mirrors of a strong-microcavity device are metal: an opaque Ag bottom mirror and a semi-transparent Al top electrode. Only wavelengths that satisfy the round-trip phase condition build a standing wave and escape, so the resonance wavelength scales with the effective cavity length and inversely with the resonance order j. As the observation angle grows, propagation inside the cavity tilts away from the normal, the round-trip phase shortens, and the resonance moves toward the blue.

The quality factor Q is the peak wavelength divided by the full width at half maximum. Q of a Fabry–Pérot cavity grows roughly in proportion to the order, so dividing by j leaves a quantity set only by mirror loss, comparable across orders. The paper reports Q/j = 24.4 ± 3.3 for all its devices.

Structure

The three devices differ only in the NPB, Alq3, and BPhen thicknesses; both metal mirrors and both injection layers are identical. Rows run from the outcoupling side to the substrate side.

Layerλ/2 deviceλ device3λ/2 device
Al semi-transparent top electrode30 nm30 nm30 nm
LiF1 nm1 nm1 nm
BPhen60 nm147 nm192 nm
Alq3 emissive layer20 nm40 nm40 nm
NPB39 nm102 nm133 nm
MoOx1 nm1 nm1 nm
Ag bottom mirror100 nm100 nm100 nm
Total organic thickness119 nm289 nm365 nm
Resonance order j123

The incidence medium is air and the transmission medium is a constant n=3.9n=3.9 standing in for the Si wafer, since the 100 nm Ag already blocks the backward path. All refractive indices come from measured dispersion in the built-in material database, with no scaling.

The paper publishes two sets of layer thicknesses for the same devices: measured values in Table 1 of the main text, and the values in Supplementary Table S1 used for the theory-versus-experiment comparison of Figure 4. This case uses the latter, because Figure 4 was drawn with it. The paper also publishes no numerical refractive-index table, stating only n=1.8n=1.8 for the organic layers inside its analytic model, so the absolute peak position carries a systematic offset that cannot be closed.

The λ/2 device, 119 nm of organic material:

The λ device, all three layers thickened to 289 nm total:

The 3λ/2 device, 365 nm total:

Optical and Emitter Settings

Only the Alq3 layer is emissive, and the emitter settings are identical across the three devices.

SettingValueSource
Emissive layerAlq3, 20 nm in the λ/2 device and 40 nm in the λ and 3λ/2 devicesSupplementary Table S1
Dipole distributionTen equally spaced dipole planes in the emissive layer, exponential exciton profile, diffusion length 3 nm, weight peaking at the hole-transport/emissive interfaceSupplementary information
Dipole orientationHorizontal to vertical 30:1, entered as a vertical fraction of 0.0323Supplementary information
Source spectrumAlq3 photoluminescence from the built-in material database, 36 points, 435.9–785.4 nm, peak 539.2 nmSubstitute input; the paper states only a 526 nm peak
DetectorEmission IntensityCounts only the far field emitted into the air above
Wavelength sampling440–780 nm, step 1 nm
Observation-angle sampling0–70°, step Covers the horizontal axis of paper Figure 4

The dipole-plane settings in the emissive layer:

The orientation field takes the vertical fraction, so a horizontal-dominant 30:1 ratio is entered as 1/31 ≈ 0.0323. Entering the reciprocal produces the opposite polarization behaviour.

The wavelength and angle ranges for Intensity:

Both the angular dispersion and the quality factor are read from this single run; no additional detector is needed.

Simulation Results and Comparison with Figure 4

The published figure, with the authors' own simulation on the left and measurement on the right:

Dahal Figure 4 — angle-resolved electroluminescence of the λ/2, λ, and 3λ/2 resonance-order devices; the left column is the authors' own TMM simulation and the right column is measurement.Dahal et al., Scientific Reports 11, 8456 (2021), Figure 4CC BY

Switching the Intensity result to Heatmap gives the same chart type as the paper: observation angle across, wavelength up, and the bright band is the cavity resonance. The λ/2 device bends the most and splits into two branches beyond 40°:

The λ device, with a narrower band:

The 3λ/2 device, with the narrowest band and the least bending:

Angular dispersion shape

Subtracting each device's own normal-direction peak from its per-angle peak leaves a pure shape curve, which is then compared against column-by-column digitization of both panels of paper Figure 4.

DeviceRMS / maximum difference vs. the paper's simulationRMS / maximum difference vs. the paper's measurement
λ/24.0 / 16.0 nm4.5 / 14.7 nm
λ4.2 / 12.3 nm2.1 / 4.6 nm
3λ/24.1 / 8.2 nm2.3 / 4.6 nm

The largest differences sit at 40–50°, where the s and p branches alternate as the global maximum. Total blue shift from 0 to 60°:

DeviceThis runPaper simulationPaper measurement
λ/229.8 nm31.1 nm34.5 nm
λ49.9 nm54.0 nm54.4 nm
3λ/237.1 nm45.3 nm39.4 nm

All three datasets show that the blue shift is not monotonic in resonance order. Angular color shift is set jointly by the normal-direction peak and the effective index, so thickening a cavity to a higher order does not automatically buy a smaller shift.

Quality factor and linewidth compression

The quality factor comes from the peak and full width at half maximum of the normal-direction spectrum, covering all fifteen cavity thicknesses reported in the paper.

Resonance order jDevicesFWHMQQ/j
1618.9–36.8 nm21–2725.6
2511.6–12.8 nm42–4822.8
347.5–8.2 nm56–6219.7

Across all devices Q/j is 23.1 ± 3.0 against the paper's 24.4 ± 3.3, a relative difference of 5.3% with overlapping 1σ intervals. The non-cavity reference device has an electroluminescence FWHM of about 100 nm; placing the same emitter in a strong microcavity compresses the emitted linewidth to 7.5–36.8 nm, more strongly at higher order.

s and p polarization splitting

At normal incidence the two polarizations are degenerate. As the angle grows, the p branch moves to the red of the s branch, forming the swallowtail shape described in the supplementary information.

Quantity20°40°60°70°
λ/2 device splitting (p peak minus s peak)+4.7 nm+17.4 nm+31.3 nm+36.5 nm
λ device splitting+2.3 nm+8.3 nm+15.2 nm+17.5 nm
3λ/2 device splitting+1.6 nm+5.6 nm+8.3 nm
λ/2 device p-to-s peak intensity ratio0.991.172.615.91

All three qualitative claims reproduce: s and p are degenerate at normal incidence, the s peak always sits on the blue side, and p strengthens at large angles. The splitting shrinks with order, matching the supplementary figure where the λ/2 swallowtail is the most pronounced. The s branch of the 3λ/2 device at 70° is too weak to locate a stable peak.

Deviation Notes

The paper does not publish the NPB, Alq3, and BPhen dispersion it actually used, stating only n=1.8n=1.8 for the organic layers inside its analytic model. Rerunning with a constant organic index of 1.80 brackets the offset:

DeviceThis run (database dispersion)Rerun with n=1.80n=1.80Paper simulation
λ/2560.2 nm574.9 nm567.5 nm
λ578.4 nm593.7 nm593.8 nm
3λ/2485.4 nm488.2 nm490.2 nm

The index set the paper actually used lies between the two runs, so the absolute normal-direction peak carries a −15.4 to +7.4 nm systematic offset that cannot be narrowed further. The angular dispersion shape and Q/j compared above are both differential quantities and are unaffected.

Layer thicknesses come from Supplementary Table S1. The source spectrum peaks 13.2 nm to the red of the 526 nm the paper states; shifting the whole spectrum by −13.2 nm and rerunning moves the peaks by only −0.22 nm on average.

The supplementary information folds electron-transport-layer surface roughness (RMS 0–8 nm) into its scattering matrix, which this model has no counterpart for; the expected consequence is a slightly narrower line here than in the paper.

Back to Case Studies

Copyright © 2026 Dreapex